BAM reference data: EDS raw data of Al-coated titania nanoparticles (JRCNM62001a and JRCNM62002a)

Description

The EDS spectra are given in the EMSA/MAS format as defined by ISO 22029:2012 Microbeam analysis — EMSA/MAS standard file format for spectral-data exchange. The exact locations of the sample areas measured with EDS are indicated in the SEM images. For further information please look at: - Radnik, J. Kersting, R., Hagenhoff, B., Bennet, F., Ciornii, D.; Nymark, P., Grafström R. and Hodoroaba, V.- D. Nanomaterials 2021, 11, 639. https://doi.org/10.3390/nano11030639, and - Radnik, Jörg. (2021). BAM reference data: XPS raw data of Al-coated titania nanoparticles (JRCNM62001a and JRCNM62002a) [Data set]. Nanomaterials. Zenodo. http://doi.org/10.5281/zenodo.4986068 Measurement conditions: The EDS analysis in the present study has been performed with a QUANTAX 400 EDS system (BRUKER, Berlin, Germany), which is equipped with an SDD (Silicon Drift-Detector) of the 10 mm2 nominal area. An excitation of 10 keV was applied for the analysis of the titania samples prepared as a thick dry powder layer on an aluminum stub, so that the substrate cannot be coexcited. The analysis  reas were selected as large as 5 x 5 µm2 on sample agglomerates of about 1eng0 µm size.

Resources

Name Format Description Link
Unknown Resource Unknown Open resource
Unknown Resource Unknown Open resource
Unknown Resource Unknown Open resource
Unknown Resource Unknown Open resource
Unknown Resource Unknown Open resource

Tags

Topics

Categories