| Name |
Format |
Description |
Link |
|
47 |
Readme for the navigation, organization, and content for the data in the figures included in the Measuring VHF Detector Linearity using a Quantum-Based Source CPEM 2024 abstract. |
https://data.nist.gov/od/ds/mds2-3155/Readme.txt |
|
8 |
Power sensor linearity measured using the VHF-JAWS source for frequencies up to 50.05 MHz. Power sensor residuals (nW), i.e., differences from a linear (mx+b) fit, and standard uncertainties (nW) vs. Expected Power (uW) for 1.5 MHz, 5.5 MHz, 15 MHz, 30 MHz, 40 MHz, and 50.05 MHz. Each power sensor residual value is derived from averaging over a set of 12 measurements within a mini-QLR. The uncertainties are the standard deviation of each measurement set. |
https://data.nist.gov/od/ds/mds2-3155/figure3_bottom_residuals_power_sensor.csv |
|
8 |
Digitizer linearity measured using the VHF-JAWS source for frequencies up to 50.05 MHz. Digitizer residuals (nW), i.e., differences from a linear (mx+b) fit, and standard uncertainties (nW) vs. Expected Power (uW) for 50 kHz, 100 kHz, 1.5 MHz, 5.5 MHz, 15 MHz, 30 MHz, 40 MHz, and 50.05 MHz. Each digitizer residual value is derived from averaging over a set of 12 measurements within a mini-QLR. The uncertainties are the standard deviation of each measurement set. |
https://data.nist.gov/od/ds/mds2-3155/figure3_top_residuals_digitizer.csv |
|
8 |
DC offset quantum locking ranges (QLRs) measured by the digitizer for frequencies up to 50.05 MHz and an expected output rms voltage of 25 mV. (top figure) RMS voltage magnitude (mV) vs. dc offset (mA) and (bottom figure) total harmonic distortion (THD) vs. dc offset (mA) collected by the digitizer for 50 kHz, 100 kHz, 1.5 MHz, 5.5 MHz, 15 MHz, 30 MHz, 40 MHz, and 50.05 MHz waveforms. In the figure, voltage measurements for the 30 MHz tone are offset by +6.5 mV for visual clarity. This offset is included in the dc_qlr_vrms column in this file. |
https://data.nist.gov/od/ds/mds2-3155/figure2_dc_qlr.csv |