Description
Background:
Head extension and excessive laryngoscope blade levering motion
(LBLM) are undesirable during airway management of trauma patients. We
hypothesized that laryngoscopy with a modified blade facilitating glottic
exposure by balloon inflation would reduce head extension and LBLM.
Patients and methods:
Seventeen elective surgery patients were enrolled. Patients lay
supine with their heads flat on a rigid board and had a rigid collar around
their necks. Laryngoscopy was performed with the modified blade and a standard
curved blade. Head extension and LBLM angles were determined upon maximal
glottic exposure and compared used paired t-tests. Laryngoscopic view
grade and oxygen saturation were also determined.
Results:
Balloon laryngoscopy resulted in less head extension and LBLM
(P <0.001). Laryngoscopic view was approximately identical with
both blades, and oxygen saturation was always above 97%.
Conclusions:
Balloon laryngoscopy reduces head extension and LBLM under
simulated cervical spine precautions.